This GRUNDIG UV5A Owner Manual is not only an instruction manual but a total functional circuit description of the Multimeter and includes circuit diagram at the end pages. It is very helpful for repairing and calibrating the instrument. It is written in two languages English and German for international support. It was very easy to repair my unit with this document. Regards, Regis Pauly, Electrical Engineer.
The owner's manual was evetything I thought it would be. I was able to print it out using both sides of the paper in quick order, saving me a search online for a used one.
Fantastic quality and impressive delivery time. Based on this manual, I recommend this service without hesitation! Five Stars for certain.
Text excerpt from page 11 (click to view)
PT-D5500U / PT-D5500E / PT-D5500UL / PT-D5500EL
[2nd page]
[3rd page]
4. CW INDEX When the color wheel is replaced, adjusts it with and buttons. a. Decrease CW INDEX setting value from the default value (289) by 50, and set it to 239. A blue-purple color mixture appears in the upper right or the upper of the screen. (When CEILING setting, appears it in the lower left or the lower of the screen.) b. Increase CW INDEX setting value by 1, and record the value where the color mixture disappears. (The recorded value is assumed to "A".) c. Increase CW INDEX setting value from the default value (289) by 50, and set it to 339. A yellow-green color mixture appears in the lower right or the lower of the screen. (When CEILING setting, appears it in the upper left or the upper of the screen.) d. Decrease CW INDEX setting value by 1, and record the value where the color mixture disappears. (The recorded value is assumed to "B".) e. Set the mean value (omission below decimal point) of "A" and "B" to the CW INDEX setting value.